Research Article| Volume 92, P244-251, November 2021

Automated interictal source localisation based on high-density EEG

  • Author Footnotes
    1 Permanent address: Epilepsy-Center Berlin-Brandenburg, Department of Neurology, Charité – Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany
    Bernd J. Vorderwülbecke
    Correspondence
    Corresponding author at: EEG and Epilepsy Unit, Epilepsy and Brain Networks Group, University Hospitals and Faculty of Medicine, University of Geneva, Rue Gabrielle-Perret-Gentil 4, 1205 Geneva, Switzerland.
    Footnotes
    1 Permanent address: Epilepsy-Center Berlin-Brandenburg, Department of Neurology, Charité – Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany
    Affiliations
    EEG and Epilepsy Unit, University Hospitals and Faculty of Medicine, University of Geneva, Rue Gabrielle-Perret-Gentil 4, 1205 Geneva, Switzerland

    Epilepsy-Center Berlin-Brandenburg, Department of Neurology, Charité – Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany
    Search for articles by this author
  • Amir G. Baroumand
    Affiliations
    Medical Image and Signal Processing Group, Department of Electronics and Information Systems, Ghent University, Corneel Heymanslaan 10, 9000 Ghent, Belgium

    Epilog NV, Vlasgaardstraat 52, 9000 Ghent, Belgium
    Search for articles by this author
  • Laurent Spinelli
    Affiliations
    EEG and Epilepsy Unit, University Hospitals and Faculty of Medicine, University of Geneva, Rue Gabrielle-Perret-Gentil 4, 1205 Geneva, Switzerland
    Search for articles by this author
  • Margitta Seeck
    Affiliations
    EEG and Epilepsy Unit, University Hospitals and Faculty of Medicine, University of Geneva, Rue Gabrielle-Perret-Gentil 4, 1205 Geneva, Switzerland
    Search for articles by this author
  • Pieter van Mierlo
    Affiliations
    Medical Image and Signal Processing Group, Department of Electronics and Information Systems, Ghent University, Corneel Heymanslaan 10, 9000 Ghent, Belgium

    Epilog NV, Vlasgaardstraat 52, 9000 Ghent, Belgium
    Search for articles by this author
  • Serge Vulliémoz
    Affiliations
    EEG and Epilepsy Unit, University Hospitals and Faculty of Medicine, University of Geneva, Rue Gabrielle-Perret-Gentil 4, 1205 Geneva, Switzerland
    Search for articles by this author
  • Author Footnotes
    1 Permanent address: Epilepsy-Center Berlin-Brandenburg, Department of Neurology, Charité – Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany
Open AccessPublished:September 30, 2021DOI:https://doi.org/10.1016/j.seizure.2021.09.020

      Highlights

      • Interictal spikes were automatically detected in 21–24 of 30 evaluated patients.
      • Compared to resection and outcome, source localisation was accurate in 55–71%.
      • Low- and high-density EEG with 25–257 channels had similar diagnostic accuracies.

      Abstract

      Purpose

      To study the accuracy of automated interictal EEG source localisation based on high-density EEG, and to compare it to low-density EEG.

      Methods

      Thirty patients operated for pharmacoresistant focal epilepsy were retrospectively examined. Twelve months after resective brain surgery, 18 were seizure-free or had ‘auras’ only, while 12 had persistence of disabling seizures. Presurgical 257-channel EEG lasting 3–20 h was down-sampled to 25, 40, and 204 channels for separate analyses. For each electrode setup, interictal spikes were detected, clustered, and averaged automatically before validation by an expert reviewer. An individual 6-layer finite difference head model and the standardised low-resolution electromagnetic tomography were used to localise the maximum source activity of the most prevalent spike. Sublobar concordance with the resected brain area was visually assessed and related to favourable vs. unfavourable postsurgical outcome.

      Results

      Depending on the EEG setup, epileptic spikes were detected in 21-24 patients (70-80%). The median number of single spikes per average was 470 (range 17–15,066). Diagnostic sensitivity of EEG source localisation was 58-75%, specificity was 50-67%, and overall accuracy was 55-71%. There were no significant differences between low- and high-density EEG setups with 25 to 257 electrodes.

      Conclusion

      Automated high-density EEG source localisation provides meaningful information in the majority of cases. With hundreds of single spikes averaged, diagnostic accuracy is similar in high- and low-density EEG. Therefore, low-density EEG may be sufficient for interictal EEG source localisation if high numbers of spikes are available.

      Keywords

      Abbreviations

      CI
      confidence interval
      ESL
      EEG source localisation
      FDM
      finite difference model
      FN
      false negative
      FP
      false positive
      IFCN
      International Federation of Clinical Neurophysiology
      ILAE
      International League Against Epilepsy
      IQR
      interquartile range
      sLORETA
      standardised low-resolution electromagnetic tomography
      OR
      odds ratio
      TN
      true negative
      TP
      true positive

      1. Introduction

      Pharmacoresistant focal epilepsy can be cured by surgical removal of the epileptogenic zone, i.e., the specific area of the brain which is indispensable for the generation of epileptic seizures [
      • Rosenow F.
      • Lüders H.
      Presurgical evaluation of epilepsy.
      ]. Prior to such surgery, potentially eligible patients undergo a thorough multimodal diagnostic workup, first, to delineate the epileptogenic zone with maximum possible precision, and second, to define functional brain areas which must remain untouched to avoid neurological complications [
      • Ryvlin P.
      • Cross J.H.
      • Rheims S.
      Epilepsy surgery in children and adults.
      ,
      • Zijlmans M.
      • Zweiphenning W.
      • van Klink N.
      Changing concepts in presurgical assessment for epilepsy surgery.
      ]. One non-invasive neurophysiological tool to identify the epileptogenic zone is EEG source localisation (ESL). Based on biophysical models, ESL plots the sources of scalp EEG activity within a 3D reconstruction of the brain. For presurgical epilepsy evaluation, ESL can be applied both to EEG seizure patterns and to interictal epileptic discharges (‘spikes’). Ictal ESL is technically demanding because EEG seizure patterns display complex spatiotemporal dynamics and are often corrupted by movement artefacts [
      • van Mierlo P.
      • Vorderwülbecke B.J.
      • Staljanssens W.
      • Seeck M.
      • Vulliemoz S.
      Ictal EEG source localization in focal epilepsy: review and future perspectives.
      ]. By contrast, spikes are rather simply configured and can easily be averaged to achieve a high signal-to-noise ratio, and they usually occur far more frequently than seizures. Therefore, interictal ESL is more widely used and more extensively validated for presurgical epilepsy evaluation than ictal ESL [
      • Mouthaan B.E.
      • Rados M.
      • Boon P.
      • Carrette E.
      • Diehl B.
      • Jung J.
      • et al.
      Diagnostic accuracy of interictal source imaging in presurgical epilepsy evaluation: a systematic review from the E-PILEPSY consortium.
      ,
      • Sharma P.
      • Seeck M.
      • Beniczky S.
      Accuracy of Interictal and Ictal Electric and Magnetic Source Imaging: a Systematic Review and Meta-Analysis.
      ].
      In the vast majority of interictal ESL studies, spikes are identified visually by expert neurophysiologists or epileptologists [
      • Abdallah C.
      • Maillard L.G.
      • Rikir E.
      • Jonas J.
      • Thiriaux A.
      • Gavaret M.
      • et al.
      Localizing value of electrical source imaging: frontal lobe, malformations of cortical development and negative MRI related epilepsies are the best candidates.
      ,
      • Foged M.T.
      • Martens T.
      • Pinborg L.H.
      • Hamrouni N.
      • Litman M.
      • Rubboli G.
      • et al.
      Diagnostic added value of electrical source imaging in presurgical evaluation of patients with epilepsy: a prospective study.
      ,
      • Plummer C.
      • Vogrin S.J.
      • Woods W.P.
      • Murphy M.A.
      • Cook M.J.
      • Liley D.T.J.
      Interictal and ictal source localization for epilepsy surgery using high-density EEG with MEG: a prospective long-term study.
      ,
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ]. Since visual identification, averaging and localisation of spikes is both user-dependent and time-consuming, an automated method of epileptic spike detection and subsequent ESL has been proposed and validated in presurgical patients with focal epilepsy [
      • Baroumand A.G.
      • van Mierlo P.
      • Strobbe G.
      • Pinborg L.H.
      • Fabricius M.
      • Rubboli G.
      • et al.
      Automated EEG source imaging: a retrospective, blinded clinical validation study.
      ,
      • van Mierlo P.
      • Strobbe G.
      • Keereman V.
      • Birot G.
      • Gadeyne S.
      • Gschwind M.
      • et al.
      Automated long-term EEG analysis to localize the epileptogenic zone.
      ]. However, the respective studies were based on conventional low-density EEG with 25–31 electrodes. High-density EEG using at least 64 electrodes has been reported to yield more accurate ESL results than low-density EEG in comparative studies [
      • Brodbeck V.
      • Spinelli L.
      • Lascano A.M.
      • Wissmeier M.
      • Vargas M.I.
      • Vulliemoz S.
      • et al.
      Electroencephalographic source imaging: a prospective study of 152 operated epileptic patients.
      ,
      • Lantz G.
      • Grave de Peralta R.
      • Spinelli L.
      • Seeck M.
      • Michel C.M
      Epileptic source localization with high density EEG: how many electrodes are needed?.
      ,
      • Sohrabpour A.
      • Lu Y.
      • Kankirawatana P.
      • Blount J.
      • Kim H.
      • He B.
      Effect of EEG electrode number on epileptic source localization in pediatric patients.
      ] although this finding could not be confirmed in a recent meta-analysis [
      • Sharma P.
      • Seeck M.
      • Beniczky S.
      Accuracy of Interictal and Ictal Electric and Magnetic Source Imaging: a Systematic Review and Meta-Analysis.
      ]. The accuracy of automated high-density ESL has not yet been assessed. Therefore, the study presented here retrospectively evaluated automated interictal ESL based on presurgical 257-channel EEG, validated by subsequent resective brain surgery and 12-month postsurgical seizure outcome. Via spatial down-sampling of the full 257-channel EEG, the accuracy of spike detection and ESL was compared between two high-density and two low-density EEG setups.

      2. Methods

      2.1 Patients and recordings

      The high-density EEG database of the EEG and Epilepsy Unit at the University Hospitals Geneva was retrospectively screened, up to 31 July 2020, for patients meeting the following criteria: (a) 257-channel EEG recording with a minimum duration of 3 h (to achieve a sufficiently high number of spikes); (b) subsequent first resective brain surgery to treat pharmacoresistant focal epilepsy; (c) presurgical high-resolution structural MRI available; (d) postsurgical brain imaging and 12-month postsurgical seizure outcome available. Following the International League Against Epilepsy (ILAE) criteria [
      • Wieser H.G.
      • Blume W.T.
      • Fish D.
      • Goldensohn E.
      • Hufnagel A.
      • King D.
      • et al.
      Proposal for a new classification of outcome with respect to epileptic seizures following epilepsy surgery.
      ], patients were classified as having favourable 12-month postsurgical outcome if they were seizure-free or had ‘auras’ only since surgery (ILAE class 1 or 2, equivalent to Engel IA and IB) or unfavourable outcome if disabling seizures persisted (ILAE class 3 or higher). Favourable postsurgical outcome indicates that the epileptogenic zone was correctly identified. Prior to surgery, for at least 3 h during daytime or overnight, high-density EEG was recorded using 257 electrodes (Philips EGI, now Magstim EGI, Eugene, Oregon, USA), unfiltered acquisition, at 500–1000 Hz sampling rate. For the head model, individual high-resolution structural T1 or MPRAGE MRI without contrast medium was used. The study was approved by the local ethics committee of the province (canton) of Geneva. Since clinical routine data were reused and individuals cannot be identified, the need for written informed consent was waived.

      2.2 EEG setups

      To allow comparison of different electrode setups representing low- and high-density EEG, the original 257-channel EEG was spatially down-sampled to three other setups (Fig. 1, supplementary figure 1). The first contains 25 channels as recommended as a minimum for clinical routine low-density EEG by the International Federation of Clinical Neurophysiology (IFCN). It represents the classical 10–20 setup with an additional three basal temporal electrodes on either side, namely F9/10, T9/10, and P9/10 [
      • Seeck M.
      • Koessler L.
      • Bast T.
      • Leijten F.
      • Michel C.
      • Baumgartner C.
      • et al.
      The standardized EEG electrode array of the IFCN.
      ]. As an example for a setup typically used in presurgical epilepsy evaluation, the second array consists of 40 channels as used in the ongoing ‘Prospective Multicenter Study on Localization Accuracy and Clinical Utility of Automated Electric Source Imaging in Presurgical Evaluation’ [
      • Beniczky S.
      Clinical Utility of Automated Electric Source Imaging in Presurgical Evaluation.
      ]. As a high-density alternative to the full 257-channel array, the third down-sampled setup contains 204 channels as used in a previous study of ours [
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ]. It does not include the electrodes F9/10 and T9/10, but the electrode row 5% more cranial is preserved (Fig. 1, supplementary figure 1).
      Fig. 1
      Fig. 1Illustration of the four different EEG setups used. Two low-density EEG setups (above) were compared to two high-density setups (below) by spatial down-sampling of the full 257-channel EEG setup (bottom right). Included channels are displayed in red, excluded channels in blue. Please note that the 257-electrode array does not perfectly match the 10–10 array on which both the 25- and the 40-channel setup are based.

      2.3 Automated EEG source localisation

      Fig. 2 gives an overview of the processing steps. De-identified raw hd-EEG and MRI data were processed by Epilog NV (Ghent, Belgium), blinded for clinical information including site of surgery and outcome. For each EEG setup, electrodes overlapping with the classical 10–10 montage were pinpointed and renamed. These were all channels of the 25- and the 40-channel setup, 76 for the 204-channel setup and 84 for the 257-channel setup (supplementary figure 2). Continuously noisy channels were visually identified and excluded from further analysis.
      Fig. 2
      Fig. 2Overview of the processing steps. The full 257-channel EEG was downsampled to 25, 40, and 204 channels for comparison. For each EEG setup, spikes were automatically detected and clustered based on their topography. The four clusters with the highest numbers of single events were further evaluated. The 257-channel EEG setup was co-registered to the structural 3D MRI and downsampled. The head was automatically segmented into six tissue layers (white matter, grey matter, cerebrospinal fluid, skull, scalp, and air) to generate an individual finite difference head model (FDM). The standardised low-resolution electromagnetic tomography (sLORETA) served as inverse solution. For each spike cluster, the maximum (crosshairs) of the source activation was visualised in the three orthogonal planes of the MRI at the onset, half-rise, and peak of the average spike. The source maximum was later compared to the postsurgical MRI.
      Using the Persyst Spike Detector P14 (Persyst, San Diego, California, USA), the EEG was band-pass filtered between 0.5 and 30 Hz. All detected events with a spike-probability higher than 0.5 were automatically selected for further analysis, and a two-second epoch centred around the peak was generated for each of them. Epochs containing possible corrupted channels (i.e., channels with a standard deviation that exceeds five times the median standard deviation of all channels) were excluded, and the remaining channels were average-referenced and baseline-corrected. Based on the topographical maximum, i.e., the channel with the highest amplitude at the peak, the single event epochs were grouped into clusters and merged if the scalp topography at the peak had a correlation higher than 0.9. Clusters with less than 15 single events were excluded. Eventually, those four clusters with the highest numbers of single events were subjected to ESL analysis.
      The patient's T1 or MPRAGE MRI sequences were used to construct an individual head model with a 1 mm cubic resolution and six different tissues: scalp, skull, cerebrospinal fluid, grey matter, white matter, and air [
      • Montes-Restrepo V.
      • Carrette E.
      • Strobbe G.
      • Gadeyne S.
      • Vandenberghe S.
      • Boon P.
      • et al.
      The Role of Skull Modeling in EEG Source Imaging for Patients with Refractory Temporal Lobe Epilepsy.
      ,
      • Strobbe G.
      • Carrette E.
      • Lopez J.D.
      • Montes Restrepo V.
      • Van Roost D.
      • Meurs A.
      • et al.
      Electrical source imaging of interictal spikes using multiple sparse volumetric priors for presurgical epileptogenic focus localization.
      ]. For tissue segmentation, the Statistical Parametric Mapping software version 12 (Wellcome Centre for Human Neuroimaging, University College London Queen Square Institute of Neurology, London, UK) automatically estimated each tissue probability map. The maps were smoothed and then combined to generate the head model, with manual adjustments if necessary. The EEG electrodes were manually co-registered to the head model. Dipoles were distributed in a regular 3-dimensional grid with 4 mm spacing throughout the grey matter, excluding thalamus and basal ganglia. The lead-fields were computed using the finite difference method as forward solution [
      • Hallez H.
      • Vanrumste B.
      • Van Hese P.
      • D'Asseler Y.
      • Lemahieu I.
      • Van de Walle R
      A finite difference method with reciprocity used to incorporate anisotropy in electroencephalogram dipole source localization.
      ]. The standardised low-resolution electromagnetic tomography (sLORETA) [
      • Pascual-Marqui R.D.
      Standardized low-resolution brain electromagnetic tomography (sLORETA): technical details.
      ] served as inverse solution. For each EEG setup, ESL was performed at the time of the onset of the average spike, the half-rise of its rising flank, and its peak (Fig. 2).

      2.4 Evaluation

      To distinguish genuine spike clusters from other EEG patterns, an EEG-certified neurologist (B. J. V., aware of clinical information) visually reviewed 2-second epochs of the 20 most representative single events per cluster in bipolar and average montage, based on 18 EEG channels for visual simplicity. Artefacts and physiological EEG patterns were discarded. Polyphasic epileptiform patterns such as polyspikes or fast rhythms were not further analysed either, because the detection algorithm chooses the peak of maximum amplitude and not the first apparent deflection, and the respective source may be subject to propagation. “Mixed” clusters that consisted of both genuine spikes and artefacts were taken as artefact clusters. Sensitivity of spike detection was defined as the proportion of genuine spike clusters amongst all clusters. Success rate was defined as the proportion of EEGs with at least one genuine spike cluster.
      For each genuine spike cluster, source maxima at the onset, half-rise, and peak of the average spike were visually compared to the postsurgical MRI. The grey matter was divided into 19 ‘sublobes’ per hemisphere [
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ] based on the Lausanne parcellation atlas [
      • Daducci A.
      • Gerhard S.
      • Griffa A.
      • Lemkaddem A.
      • Cammoun L.
      • Gigandet X.
      • et al.
      The connectome mapper: an open-source processing pipeline to map connectomes with MRI.
      ]. Two subcortical ‘sublobes’ belonging to the atlas were not included in the source space (thalamus and basal ganglia; supplementary table S1). Source maxima were compared to sublobe(s), lobe(s) and hemisphere affected by resective surgery, and to the borders of the resection itself. Sublobar concordance at the most prevalent genuine spike's half-rise was the study's primary outcome parameter. Secondary outcome parameters were the source maximum being inside the resected area, sublobar concordance at spike onset and peak, and sublobar concordance using other spike clusters than the most prevalent one.
      Concordance in patients with favourable outcome (ILAE 1-2) was considered true positive (TP), and discordance in patients with unfavourable outcome (ILAE ≥ 3) was considered true negative (TN). By analogy, sublobar concordance in case of unfavourable outcome was considered false positive (FP), and discordance in favourable outcome was considered false negative (FN). Sensitivity was calculated as TP / (TP+FN), and specificity as TN / (TN+FP). Please note that TP and sensitivity are more reliably defined than TN and specificity [
      • Rikir E.
      • Koessler L.
      • Ramantani G.
      • Maillard L.G.
      Added value and limitations of electrical source localization.
      ]. Overall accuracy was calculated as (TP+TN) / (TP+TN+FP+FN). The diagnostic odds ratio (OR) of favourable postsurgical outcome in case of concordant ESL results was calculated as (TP*TN) / (FP*FN). The 95%-confidence interval (95%-CI) of the parameters sensitivity, specificity, and overall accuracy was calculated as. parameter±((parameter)*(1parameter)samplesize1.96*) The 95%-CI of the diagnostic OR was calculated as e(ln(OR)±1.96*)1TP+1TN+1FP+1FN [
      • Baroumand A.G.
      • van Mierlo P.
      • Strobbe G.
      • Pinborg L.H.
      • Fabricius M.
      • Rubboli G.
      • et al.
      Automated EEG source imaging: a retrospective, blinded clinical validation study.
      ]. These definitions were specified before analysing the data.

      2.5 Statistical analyses

      Statistical analyses were performed using SPSS 25 (IBM, Armonk, US-NY). Data are given as percent or as median and interquartile range (IQR). Dichotomous data from more than two dependant samples were subjected to Cochran Q tests; nominal (including dichotomous) data from independent samples were subjected to χ² tests. Ordinal or continuous data from two independent samples were subjected to Mann-Whitney U tests. Continuous data from > 2 dependant samples were subjected to Friedman's tests, followed by pairwise Wilcoxon tests. Correlations of continuous data were compared using Spearman's ρ. All tests were two-sided. p < 0.05 was considered statistically significant. Repetitive pairwise testing of the same parameter was corrected for using the Benjamini-Hochberg procedure [
      • Hemmerich W.
      Rechner zur Adjustierung des α-Niveaus.
      ]. However, to not inflate the risk of type-II errors, the testing of multiple parameters was not corrected for. Besides favourable vs. unfavourable outcome, no further patient subgroup analyses (e.g., temporal vs. extratemporal cases, MRI-positive vs. MRI-negative cases) were performed because of the low numbers of patients.

      3. Results

      3.1 Patient cohort

      A total of 357 hd-EEG recordings dating from 2008 to 2020 were screened, of which 47 were recorded with 257 channels and had a duration of at least 3 h. Eight patients did not undergo resective surgery, three had previous brain surgery, and one had a skull defect. In three patients, the EEG files were irreparably corrupted, and another two had relevant clinical data missing. With these 17 cases excluded, 30 were eventually analysed.
      Table 1 gives an overview of the study cohort; individual details can be found in table S2. Twelve months after surgery, 17 patients were seizure-free (ILAE 1), and one had aware non-motor seizures (‘auras’) only (ILAE 2). These 18 were counted as having favourable postsurgical seizure outcome. The remaining 12 patients had persisting disabling seizures and were counted as unfavourable outcome (ILAE 3–5).
      Table 1Demographic and clinical data of the study cohort. Data are given as n (column%) or as median [interquartile range]. Favourable outcome implies no seizures or ‘auras’ only 12 months after surgery (ILAE, class 1–2) while unfavourable outcome indicates persistence of disabling seizures (ILAE 3–5). Differences between patients with favourable and unfavourable outcome were not statistically significant (p > 0.05; two-sided Mann-Whitney U tests or χ² tests, as appropriate). TLE, temporal lobe epilepsy; ETLE, extratemporal epilepsy; ILAE, International League Against Epilepsy.
      All patientsFavourable outcome (ILAE 1-2)Unfavourable outcome (ILAE 3-5)
      Group size301812
      SexFemale20 (67%)11 (61%)9 (75%)
      Male10 (33%)7 (39%)3 (25%)
      Age at onset (years)16 [5–25]16 [4–25]17 [5–25]
      Age at surgery (years)29 [19–39]28 [15–39]29 [20–43]
      LocalisationTLE18 (60%)13 (72%)5 (42%)
      ETLE12 (40%)5 (28%)7 (58%)
      HemisphereLeft15 (50%)9 (50%)6 (50%)
      Right15 (50%)9 (50%)6 (50%)
      Structural MRILesional26 (87%)17 (94%)9 (75%)
      Non-lesional4 (13%)1 (6%)3 (25%)
      Duration of EEG (hours)15 [9–16]15 [6–17]15 [12–16]

      3.2 Spike detection

      The median duration of hd-EEG recordings was 15 h and 15 min (range 3 h and 20 min to 20 h and 40 min). The median proportion of continuously corrupted channels ranged from 0% (IQR 0–4%) in the 25-channel setup to 2.7% (IQR 1.2–5.8%) in the 257-channel setup (corrected p > 0.07; Friedman's test). Except for two setups in different patients, the automated spike detection found four clusters of supposed epileptic spikes with at least 15 single events in every EEG setup and patient. Following expert review, 38% of these were genuine interictal epileptic discharges, i.e., spikes or sharp waves. Another 6% were epileptic polyspikes or fast rhythms which were not considered further because of their polyphasic configuration limiting ESL, 11% were classified as physiological EEG patterns such as wicket spikes or K complexes, and 45% were classified as technical or biological artefacts like ECG or eyeblink artefacts. Detection accuracy did not significantly differ across EEG setups (p = 0.41; χ² test; Fig. 3).
      Fig. 3
      Fig. 3Accuracy of automated spike detection depending on the electrode setup, across all discharge clusters. Following visual inspection of 20 exemplary 2-second epochs centred around the peak, the automatically detected putative spike clusters (4 per patient and setup) were grouped into genuine interictal epileptic discharges (spikes), polyphasic epileptiform discharges, physiological EEG patterns, and artefacts. Differences between the setups were not statistically significant (p = 0.41; χ² test).
      The average detection sensitivity decreased from 57% in the most prevalent cluster to 23% in the fourth most prevalent cluster (Fig. 4A). The average success rate to detect at least one genuine spike cluster per EEG study increased from 57%, when considering the most prevalent cluster only, to 73%, when considering all four most prevalent clusters (Fig. 4B). Depending on the EEG setup, between 21 and 24 out of all 30 patients had at least one genuine spike cluster amongst the four most prevalent clusters. Of these patients, 12–14 had a favourable outcome and 9–10 had an unfavourable outcome. Four patients had no genuine spike cluster detected with any EEG setup. One of these four patients had polyspikes only; two others were known from clinical evaluation to have genuine spikes which were not identified by the methodology used here despite more than 15 h of EEG recording; the fourth patient was known to have very rare spikes. The median number of single events per genuine spike cluster was 470 (IQR 106–1340, range 17–15,066) with the number of single spikes correlating with the duration of the EEG recording (Spearman's ρ = 0.62; p < 0.001).
      Fig. 4
      Fig. 4Sensitivity and success rate of automated spike detection depending on the discharge clusters considered. A, proportion of genuine spikes in the first, second, third, and fourth most prevalent discharge cluster. B, success rate of automated spike detection to detect at least one genuine spike cluster depending on the number of the most prevalent discharge clusters included. *, corrected p < 0.05 in comparison to 1st discharge cluster with 40 channels (grey) or 257 channels (black; χ² tests).

      3.3 Source localisation

      For each patient and EEG setup, the genuine spike cluster with the highest number of single discharges was evaluated to assess ESL accuracy. An overview of the levels of concordance (sublobar, lobar, hemispheric, or contralateral) between ESL maximum and the resected brain area in patients with favourable vs. unfavourable postsurgical seizure outcome is given in supplementary figure 2. Sublobar concordance, the study's main outcome parameter, was achieved in 58–75% of cases with favourable postsurgical seizures outcome as compared to 33–50% of cases with unfavourable seizure outcome (corrected p > 0.10). The respective sensitivities, specificities, and overall diagnostic accuracies of ESL achieved with the different EEG setups are detailed in Fig. 5A, together with their 95%-CIs. Across all setups, the average sensitivity was 68%, the average specificity was 61%, and the average diagnostic accuracy was 65%. There were no statistically significant differences between the electrode setups (0.69 < p < 0.86). The diagnostic OR ranged from 1.4 (95%-CI 1.1–9.0) with the 204-channel setup to 6.0 (5.1–46.3) with 257 channels.
      Fig. 5
      Fig. 5Diagnostic sensitivity, specificity, and overall accuracy of EEG source localisation. A, parameters based on sublobar concordance between source maximum and resected brain area at the most prevalent spike average's half-rise. B, parameters based on the source maximum located inside the resected brain area at the most prevalent spike average's half-rise. Error bars indicate 95%-confidence intervals. Differences between the electrode setups were not statistically significant (p > 0.69; χ² tests).
      Across the time course of the spike, the average sensitivity based on sublobar concordance increased from 50% at onset to 72% at the peak, while the average specificity decreased from 65% to 50% (supplementary figure 3). The average overall accuracy was lower both at the spike's onset (57%) and at the peak (62%) than at its half-rise (65%). However, differences across the different time points were not statistically significant. All diagnostic parameters together with their 95%-CIs are detailed in supplementary table S3. The number of single spikes per average spike was similar in accurate (TP or TN) vs. non-accurate (FP or FN) tests (p = 1; corrected Mann-Whitney U tests).
      In a second, more rigorous validation approach, localisation was considered positive only with the source maximum being located inside the resected zone. In that case, average ESL sensitivity was 60% and average specificity was 74%, together with an average overall accuracy of 66% (Fig. 5B).
      A ‘second most prevalent’ genuine spike cluster was detected in 54 setup analyses of 18 patients. Here, based on sublobar concordance of ESL maximum and the resected brain area at the spike average's half rise, average sensitivity, specificity, and overall accuracy were 48%, 61%, and 54% (supplementary figure 4). With concordance defined by the source maximum being located inside the resected brain area, average sensitivity, specificity, and overall accuracy based on the second most prevalent genuine spike were 22%, 61%, and 39%. A ‘third’ and ‘fourth most prevalent’ genuine spike cluster were detected in 29 and 10 setup analyses, respectively. These were not further analysed due to the low numbers.
      In all these analyses, diagnostic parameters were related to the 21–24 cases with successfully detected genuine spikes in the sense of a ‘per-protocol’ analysis. If parameters were related to all 30 patients in the sense of an ‘intention-to-examine’ analysis, and tests that could not be performed due to missing genuine spikes were counted as negative results, the average sensitivity based on sublobar concordance at the most prevalent spike's half-rise was 47%, alongside with an average specificity of 69% and an average overall accuracy of 56%.

      4. Discussion

      With this study, we show that automated spike detection can identify genuine epileptic spikes in two thirds of cases both from high- and low-density EEG. We further demonstrate that if spikes are successfully detected, ESL localises the epileptogenic zone correctly in two thirds of cases, again irrespective of high- or low-density EEG.

      4.1 Spike detection

      On a cluster level, the sensitivity of automated spike detection ranged from 57% in the most prevalent cluster to 23% in the fourth most prevalent cluster, with an average of 38%. For comparison, on a single spike level and in a different study cohort, the same spike detector as used here had a sensitivity of 44% [
      • Scheuer M.L.
      • Bagic A.
      • Wilson S.B.
      Spike detection: inter-reader agreement and a statistical Turing test on a large data set.
      ], while other algorithms had similar sensitivities of 44–47% [
      • Ganglberger W.
      • Gritsch G.
      • Hartmann M.
      • Fürbass F.
      • Perko H.
      • Skupch A.
      • et al.
      A Comparison of Rule-Based and Machine Learning Methods for Classification of Spikes in EEG.
      ,
      • Tjepkema-Cloostermans M.C.
      • de Carvalho R.C.V.
      • van Putten M.
      Deep learning for detection of focal epileptiform discharges from scalp EEG recordings.
      ]. Experienced human EEG readers were found to have pairwise sensitivities of 40–52% for each other's spike markings [
      • Scheuer M.L.
      • Bagic A.
      • Wilson S.B.
      Spike detection: inter-reader agreement and a statistical Turing test on a large data set.
      ]. Most recently, an artificial intelligence-based algorithm was reported to detect single epileptic spikes with a sensitivity of 82% [
      • Fürbass F.
      • Kural M.A.
      • Gritsch G.
      • Hartmann M.
      • Kluge T.
      • Beniczky S.
      An artificial intelligence-based EEG algorithm for detection of epileptiform EEG discharges: validation against the diagnostic gold standard.
      ]. In our approach, assessing the sensitivity of spike detection on a single discharge level was not possible. However, the low sensitivity of 38% on a cluster level is at least partly due to the liberal settings of the spike detection algorithm with a pre-specified spike probability of 0.5.
      On an EEG analysis level, this liberal approach together with the use of up to four discharge clusters led to successful spike detection in 70–80% of patients. For comparison, 93% of the patients were known from clinical evaluation to have epileptic spikes in their EEG. This means that spikes were missed in 13–23% of evaluated patients, most likely because our approach was based on evaluating the four most prevalent EEG patterns only, and infrequent spikes could be missed. On the other hand, in comparison to manual marking and averaging of spikes, visual validation of automatically detected spike clusters based on exemplary epochs was time-efficient and convenient.

      4.2 Source localisation

      Validated by sublobar concordance with the resected brain area and 12-month surgical seizure outcome, we found that automated ESL had diagnostic sensitivities of 58–75% and overall accuracies of 54–71%. This is in the range of our earlier ESL studies [
      • Baroumand A.G.
      • van Mierlo P.
      • Strobbe G.
      • Pinborg L.H.
      • Fabricius M.
      • Rubboli G.
      • et al.
      Automated EEG source imaging: a retrospective, blinded clinical validation study.
      ,
      • van Mierlo P.
      • Strobbe G.
      • Keereman V.
      • Birot G.
      • Gadeyne S.
      • Gschwind M.
      • et al.
      Automated long-term EEG analysis to localize the epileptogenic zone.
      ,
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ] and those of others, e.g., [
      • Feng R.
      • Hu J.
      • Pan L.
      • Wu J.
      • Lang L.
      • Jiang S.
      • et al.
      Application of 256-channel dense array electroencephalographic source imaging in presurgical workup of temporal lobe epilepsy.
      ,
      • Sharma P.
      • Scherg M.
      • Pinborg L.H.
      • Fabricius M.
      • Rubboli G.
      • Pedersen B.
      • et al.
      Ictal and interictal electric source imaging in pre-surgical evaluation: a prospective study.
      ]. We previously published two articles on automated ESL based on low-density EEG. In 32 operated patients (24 of those with favourable outcome), ESL at the peak of the spike average led to a sensitivity of 70% and an overall diagnostic accuracy of 77% [
      • van Mierlo P.
      • Strobbe G.
      • Keereman V.
      • Birot G.
      • Gadeyne S.
      • Gschwind M.
      • et al.
      Automated long-term EEG analysis to localize the epileptogenic zone.
      ]. In a second study on 41 operated patients (25 with favourable outcome), ‘semi-automated’ ESL provided a sensitivity of 88% alongside with an overall accuracy of 78% [
      • Baroumand A.G.
      • van Mierlo P.
      • Strobbe G.
      • Pinborg L.H.
      • Fabricius M.
      • Rubboli G.
      • et al.
      Automated EEG source imaging: a retrospective, blinded clinical validation study.
      ]. ‘Semi-automated’ meant that the reviewer could choose other spike clusters and other time points than those suggested by default, depending on the clinical context.
      According to a meta-analysis including 515 patients from 19 studies published before June 2018, the reported sensitivity of interictal ESL is 81% (95%-CI 76–95%) and the reported overall accuracy 74% (70–78%) [
      • Sharma P.
      • Seeck M.
      • Beniczky S.
      Accuracy of Interictal and Ictal Electric and Magnetic Source Imaging: a Systematic Review and Meta-Analysis.
      ]. Sensitivities and accuracies achieved in the current study are slightly lower, but within the same range. Interestingly, in our study, ESL based on high-density EEG with 204–257 channels did not perform better than low-density EEG with 25–40 channels, whereas earlier comparative studies using 13–55 spikes per patient reported high-density EEG to lead to higher ESL accuracy than low-density EEG [
      • Brodbeck V.
      • Spinelli L.
      • Lascano A.M.
      • Wissmeier M.
      • Vargas M.I.
      • Vulliemoz S.
      • et al.
      Electroencephalographic source imaging: a prospective study of 152 operated epileptic patients.
      ,
      • Foged M.T.
      • Martens T.
      • Pinborg L.H.
      • Hamrouni N.
      • Litman M.
      • Rubboli G.
      • et al.
      Diagnostic added value of electrical source imaging in presurgical evaluation of patients with epilepsy: a prospective study.
      ,
      • Lantz G.
      • Grave de Peralta R.
      • Spinelli L.
      • Seeck M.
      • Michel C.M
      Epileptic source localization with high density EEG: how many electrodes are needed?.
      ]. Nevertheless, in the above-mentioned meta-analysis, high- and low-density EEG had virtually the same diagnostic sensitivities and accuracies [
      • Sharma P.
      • Seeck M.
      • Beniczky S.
      Accuracy of Interictal and Ictal Electric and Magnetic Source Imaging: a Systematic Review and Meta-Analysis.
      ]. We speculate that for localising average spikes with a smooth spatial distribution, based on a high number of single events, low-density EEG is sufficient. On the other hand, high-density EEG may be beneficial for localising spike averages with a low signal-to-noise ratio (due to low numbers of single spikes, low amplitude, or background noise) or single spikes, for source connectivity analyses, and for visual spike analyses in the sensor space.
      In another recent study on interictal ESL, we found that spatial down-sampling of the 257-channel array to 204 channels by removing the inferior channels over cheek and neck improved ESL accuracy [
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ]. Surprisingly, we could not replicate this finding in the current study where the 204-channel setup tended to perform worse than the full 257-channel setup. This discrepancy may result from the use of different head models: Other than a simplified realistic 3-shell head model as in the previous study, a realistic six-layer finite difference model was used in this study, which particularly accounts for the anatomy of the skull base. Therefore, inclusion of inferior channels might reduce ESL accuracy in the simplified head model but not in the more sophisticated model.
      In our study, ESL at the half-rise of the average spike was more accurate than at its onset or peak. This is in line with previous findings of ourselves [
      • Vorderwülbecke B.J.
      • Carboni M.
      • Tourbier S.
      • Brunet D.
      • Seeber M.
      • Spinelli L.
      • et al.
      High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
      ], but higher accuracies at very early time points [
      • Plummer C.
      • Vogrin S.J.
      • Woods W.P.
      • Murphy M.A.
      • Cook M.J.
      • Liley D.T.J.
      Interictal and ictal source localization for epilepsy surgery using high-density EEG with MEG: a prospective long-term study.
      ] or the peak [
      • van Mierlo P.
      • Strobbe G.
      • Keereman V.
      • Birot G.
      • Gadeyne S.
      • Gschwind M.
      • et al.
      Automated long-term EEG analysis to localize the epileptogenic zone.
      ] have also been found. Interictal spike sources at late time points may be influenced by propagation [
      • Bast T.
      • Boppel T.
      • Rupp A.
      • Harting I.
      • Hoechstetter K.
      • Fauser S.
      • et al.
      Noninvasive source localization of interictal EEG spikes: effects of signal-to-noise ratio and averaging.
      ,
      • Lantz G.
      • Spinelli L.
      • Seeck M.
      • de Peralta Menendez R.G.
      • Sottas C.C.
      • Michel C.M.
      Propagation of interictal epileptiform activity can lead to erroneous source localizations: a 128-channel EEG mapping study.
      ,
      • Merlet I.
      • Paetau R.
      • Garcia-Larrea L.
      • Uutela K.
      • Granstrom M.L.
      • Mauguiere F.
      Apparent asynchrony between interictal electric and magnetic spikes.
      ,
      • Plummer C.
      • Wagner M.
      • Fuchs M.
      • Harvey A.S.
      • Cook M.J.
      Dipole versus distributed EEG source localization for single versus averaged spikes in focal epilepsy.
      ], but the signal-to-noise ratio at a spike's onset is much lower than at the peak [
      • Bast T.
      • Boppel T.
      • Rupp A.
      • Harting I.
      • Hoechstetter K.
      • Fauser S.
      • et al.
      Noninvasive source localization of interictal EEG spikes: effects of signal-to-noise ratio and averaging.
      ,
      • Ray A.
      • Tao J.X.
      • Hawes-Ebersole S.M.
      • Ebersole J.S.
      Localizing value of scalp EEG spikes: a simultaneous scalp and intracranial study.
      ,
      • Wennberg R.
      • Cheyne D.
      EEG source imaging of anterior temporal lobe spikes: validity and reliability.
      ]. Therefore, in most cases, the half-rise of the spike can be the ‘happy medium’ to balance the effects of propagation and signal-to-noise ratio.

      4.3 Limitations and outlook

      As a limitation of this study, visual review of the automatically labelled discharges was performed by a single EEG expert and based on 20 examples per cluster. Therefore, post-hoc inclusion or exclusion of single discharges was not possible. The study results come with large 95%-confidence intervals, and most differences between the groups were not statistically significant. This is because the study cohort was limited to 30 patients, due to the strict inclusion criteria. As another limitation, the vast majority of patients had a potentially epileptogenic lesion on MRI. Concordance of interictal ESL and MRI results is a good predictor of favourable surgical outcome [
      • Lascano A.M.
      • Perneger T.
      • Vulliemoz S.
      • Spinelli L.
      • Garibotto V.
      • Korff C.M.
      • et al.
      Yield of MRI, high-density electric source imaging (HD-ESI), SPECT and PET in epilepsy surgery candidates.
      ] which underlines the usefulness of ESL also in MRI-positive cases. However, automated ESL needs to be assessed in a larger population of MRI-negative cases. For the future, prospective, adequately powered and, for that purpose, multicentre studies are necessary to better delineate the usefulness of high-density EEG for interictal source localisation.

      5. Conclusion

      Automated spike detection from high-density EEG captures interictal epileptic spikes in three out of four cases. Expert review of the putative spike clusters is essential to distinguish genuine spikes from other EEG patterns. The automated approach appears particularly useful in patients with frequent spikes. Here, diagnostic accuracies of automated high- and low-density ESL appear comparable.

      CRediT authorship contribution statement

      Bernd J. Vorderwülbecke: Conceptualization, Investigation, Formal analysis, Visualization, Writing – original draft, Writing – review & editing. Amir G. Baroumand: Investigation, Formal analysis, Validation, Writing – original draft, Writing – review & editing. Laurent Spinelli: Funding acquisition, Data curation, Writing – review & editing. Margitta Seeck: Resources, Supervision, Writing – review & editing. Pieter van Mierlo: Conceptualization, Resources, Supervision, Writing – review & editing. Serge Vulliémoz: Funding acquisition, Resources, Supervision, Writing – review & editing.

      Declaration of Competing Interest

      A. G. Baroumand is employee of epilogue NV (Ghent, BE); P. van Mierlo is co-founder and shareholder of epilogue NV; M. Seeck and S. Vulliémoz are shareholders and advisors of epilogue NV. B. J. Vorderwülbecke and L. Spinelli have no conflicts of interest to report.

      Acknowledgements

      This study was funded by the German Research Foundation [grant DFG 422589384 to B. J. Vorderwülbecke], the Swiss National Science Foundation [grants SNSF 163398 to M. Seeck, CRSII5 180365 to M. Seeck and P. van Mierlo, CRSII5 170873 to P. van Mierlo and S. Vulliémoz, SNSF 192749 to S. Vulliémoz], and the Fondation Privé HUG [grant RC1-23 to M. Seeck]. The funders were not involved in the study design, in the collection, analysis and interpretation of data, in the writing of the report, or in the decision to submit the article for publication.

      Appendix. Supplementary materials

      References

        • Abdallah C.
        • Maillard L.G.
        • Rikir E.
        • Jonas J.
        • Thiriaux A.
        • Gavaret M.
        • et al.
        Localizing value of electrical source imaging: frontal lobe, malformations of cortical development and negative MRI related epilepsies are the best candidates.
        Neuroimage Clin. 2017; 16 ([doi];S2213-1582(17)30199-7 [pii]): 319-329https://doi.org/10.1016/j.nicl.2017.08.009
        • Baroumand A.G.
        • van Mierlo P.
        • Strobbe G.
        • Pinborg L.H.
        • Fabricius M.
        • Rubboli G.
        • et al.
        Automated EEG source imaging: a retrospective, blinded clinical validation study.
        Clin Neurophysiol. 2018; https://doi.org/10.1016/j.clinph.2018.09.015
        • Bast T.
        • Boppel T.
        • Rupp A.
        • Harting I.
        • Hoechstetter K.
        • Fauser S.
        • et al.
        Noninvasive source localization of interictal EEG spikes: effects of signal-to-noise ratio and averaging.
        J Clin Neurophysiol. 2006; 23: 487-497https://doi.org/10.1097/01.wnp.0000232208.14060.c7
        • Beniczky S.
        Clinical Utility of Automated Electric Source Imaging in Presurgical Evaluation.
        (PROMAESIS). 2020; ([Accessed 09.06.2021 2021])
        • Brodbeck V.
        • Spinelli L.
        • Lascano A.M.
        • Wissmeier M.
        • Vargas M.I.
        • Vulliemoz S.
        • et al.
        Electroencephalographic source imaging: a prospective study of 152 operated epileptic patients.
        Brain. 2011; 134 (Pt 10awr243 [pii];10.1093/brain/awr243 [doi]): 2887-2897
        • Daducci A.
        • Gerhard S.
        • Griffa A.
        • Lemkaddem A.
        • Cammoun L.
        • Gigandet X.
        • et al.
        The connectome mapper: an open-source processing pipeline to map connectomes with MRI.
        PLoS ONE. 2012; 7: e48121https://doi.org/10.1371/journal.pone.0048121
        • Feng R.
        • Hu J.
        • Pan L.
        • Wu J.
        • Lang L.
        • Jiang S.
        • et al.
        Application of 256-channel dense array electroencephalographic source imaging in presurgical workup of temporal lobe epilepsy.
        Clin Neurophysiol. 2016; 127: 108-116https://doi.org/10.1016/j.clinph.2015.03.009
        • Foged M.T.
        • Martens T.
        • Pinborg L.H.
        • Hamrouni N.
        • Litman M.
        • Rubboli G.
        • et al.
        Diagnostic added value of electrical source imaging in presurgical evaluation of patients with epilepsy: a prospective study.
        Clin Neurophysiol. 2020; 131: 324-329https://doi.org/10.1016/j.clinph.2019.07.031
        • Fürbass F.
        • Kural M.A.
        • Gritsch G.
        • Hartmann M.
        • Kluge T.
        • Beniczky S.
        An artificial intelligence-based EEG algorithm for detection of epileptiform EEG discharges: validation against the diagnostic gold standard.
        Clin Neurophysiol. 2020; 131: 1174-1179https://doi.org/10.1016/j.clinph.2020.02.032
        • Ganglberger W.
        • Gritsch G.
        • Hartmann M.
        • Fürbass F.
        • Perko H.
        • Skupch A.
        • et al.
        A Comparison of Rule-Based and Machine Learning Methods for Classification of Spikes in EEG.
        J Commun. 2017; 12: 589-595https://doi.org/10.12720/jcm.12.10.589-595
        • Hallez H.
        • Vanrumste B.
        • Van Hese P.
        • D'Asseler Y.
        • Lemahieu I.
        • Van de Walle R
        A finite difference method with reciprocity used to incorporate anisotropy in electroencephalogram dipole source localization.
        Phys Med Biol. 2005; 50: 3787-3806https://doi.org/10.1088/0031-9155/50/16/009
        • Hemmerich W.
        Rechner zur Adjustierung des α-Niveaus.
        StatistikGuru. 2016; ([Accessed 21.07.2020)
        • Lantz G.
        • Grave de Peralta R.
        • Spinelli L.
        • Seeck M.
        • Michel C.M
        Epileptic source localization with high density EEG: how many electrodes are needed?.
        Clin Neurophysiol. 2003; 114 (a): 63-69https://doi.org/10.1016/s1388-2457(02)00337-1
        • Lantz G.
        • Spinelli L.
        • Seeck M.
        • de Peralta Menendez R.G.
        • Sottas C.C.
        • Michel C.M.
        Propagation of interictal epileptiform activity can lead to erroneous source localizations: a 128-channel EEG mapping study.
        J Clin Neurophysiol. 2003; 20 (b): 311-319
        • Lascano A.M.
        • Perneger T.
        • Vulliemoz S.
        • Spinelli L.
        • Garibotto V.
        • Korff C.M.
        • et al.
        Yield of MRI, high-density electric source imaging (HD-ESI), SPECT and PET in epilepsy surgery candidates.
        Clin Neurophysiol. 2016; 127 (S1388-2457(15)00314-4 [pii];10.1016/j.clinph.2015.03.025 [doi]): 150-155
        • Merlet I.
        • Paetau R.
        • Garcia-Larrea L.
        • Uutela K.
        • Granstrom M.L.
        • Mauguiere F.
        Apparent asynchrony between interictal electric and magnetic spikes.
        Neuroreport. 1997; 8: 1071-1076https://doi.org/10.1097/00001756-199703240-00002
        • Montes-Restrepo V.
        • Carrette E.
        • Strobbe G.
        • Gadeyne S.
        • Vandenberghe S.
        • Boon P.
        • et al.
        The Role of Skull Modeling in EEG Source Imaging for Patients with Refractory Temporal Lobe Epilepsy.
        Brain Topogr. 2016; 29: 572-589https://doi.org/10.1007/s10548-016-0482-6
        • Mouthaan B.E.
        • Rados M.
        • Boon P.
        • Carrette E.
        • Diehl B.
        • Jung J.
        • et al.
        Diagnostic accuracy of interictal source imaging in presurgical epilepsy evaluation: a systematic review from the E-PILEPSY consortium.
        Clin Neurophysiol. 2019; 130: 845-855
        • Pascual-Marqui R.D.
        Standardized low-resolution brain electromagnetic tomography (sLORETA): technical details.
        Methods Find Exp Clin Pharmacol. 2002; 24 (Suppl D): 5-12
        • Plummer C.
        • Vogrin S.J.
        • Woods W.P.
        • Murphy M.A.
        • Cook M.J.
        • Liley D.T.J.
        Interictal and ictal source localization for epilepsy surgery using high-density EEG with MEG: a prospective long-term study.
        Brain. 2019; 142: 932-951
        • Plummer C.
        • Wagner M.
        • Fuchs M.
        • Harvey A.S.
        • Cook M.J.
        Dipole versus distributed EEG source localization for single versus averaged spikes in focal epilepsy.
        J Clin Neurophysiol. 2010; 27: 141-162https://doi.org/10.1097/WNP.0b013e3181dd5004
        • Ray A.
        • Tao J.X.
        • Hawes-Ebersole S.M.
        • Ebersole J.S.
        Localizing value of scalp EEG spikes: a simultaneous scalp and intracranial study.
        Clin Neurophysiol. 2007; 118: 69-79https://doi.org/10.1016/j.clinph.2006.09.010
        • Rikir E.
        • Koessler L.
        • Ramantani G.
        • Maillard L.G.
        Added value and limitations of electrical source localization.
        Epilepsia. 2017; 58: 174-175https://doi.org/10.1111/epi.13643
        • Rosenow F.
        • Lüders H.
        Presurgical evaluation of epilepsy.
        Brain. 2001; 124: 1683-1700
        • Ryvlin P.
        • Cross J.H.
        • Rheims S.
        Epilepsy surgery in children and adults.
        Lancet Neurol. 2014; 13: 1114-1126https://doi.org/10.1016/S1474-4422(14)70156-5
        • Scheuer M.L.
        • Bagic A.
        • Wilson S.B.
        Spike detection: inter-reader agreement and a statistical Turing test on a large data set.
        Clin Neurophysiol. 2017; 128: 243-250https://doi.org/10.1016/j.clinph.2016.11.005
        • Seeck M.
        • Koessler L.
        • Bast T.
        • Leijten F.
        • Michel C.
        • Baumgartner C.
        • et al.
        The standardized EEG electrode array of the IFCN.
        Clin Neurophysiol. 2017; 128: 2070-2077
        • Sharma P.
        • Scherg M.
        • Pinborg L.H.
        • Fabricius M.
        • Rubboli G.
        • Pedersen B.
        • et al.
        Ictal and interictal electric source imaging in pre-surgical evaluation: a prospective study.
        Eur J Neurol. 2018; 25: 1154-1160https://doi.org/10.1111/ene.13676
        • Sharma P.
        • Seeck M.
        • Beniczky S.
        Accuracy of Interictal and Ictal Electric and Magnetic Source Imaging: a Systematic Review and Meta-Analysis.
        Front Neurol. 2019; 10: 1250https://doi.org/10.3389/fneur.2019.01250
        • Sohrabpour A.
        • Lu Y.
        • Kankirawatana P.
        • Blount J.
        • Kim H.
        • He B.
        Effect of EEG electrode number on epileptic source localization in pediatric patients.
        Clin Neurophysiol. 2015; 126: 472-480https://doi.org/10.1016/j.clinph.2014.05.038
        • Strobbe G.
        • Carrette E.
        • Lopez J.D.
        • Montes Restrepo V.
        • Van Roost D.
        • Meurs A.
        • et al.
        Electrical source imaging of interictal spikes using multiple sparse volumetric priors for presurgical epileptogenic focus localization.
        Neuroimage Clin. 2016; 11: 252-263https://doi.org/10.1016/j.nicl.2016.01.017
        • Tjepkema-Cloostermans M.C.
        • de Carvalho R.C.V.
        • van Putten M.
        Deep learning for detection of focal epileptiform discharges from scalp EEG recordings.
        Clin Neurophysiol. 2018; 129: 2191-2196https://doi.org/10.1016/j.clinph.2018.06.024
        • van Mierlo P.
        • Strobbe G.
        • Keereman V.
        • Birot G.
        • Gadeyne S.
        • Gschwind M.
        • et al.
        Automated long-term EEG analysis to localize the epileptogenic zone.
        Epilepsia Open. 2017; 2: 322-333https://doi.org/10.1002/epi4.12066
        • van Mierlo P.
        • Vorderwülbecke B.J.
        • Staljanssens W.
        • Seeck M.
        • Vulliemoz S.
        Ictal EEG source localization in focal epilepsy: review and future perspectives.
        Clin Neurophysiol. 2020; 131: 2600-2616https://doi.org/10.1016/j.clinph.2020.08.001
        • Vorderwülbecke B.J.
        • Carboni M.
        • Tourbier S.
        • Brunet D.
        • Seeber M.
        • Spinelli L.
        • et al.
        High-density Electric Source Imaging of interictal epileptic discharges: how many electrodes and which time point?.
        Clin Neurophysiol. 2020; 131: 2795-2803https://doi.org/10.1016/j.clinph.2020.09.018
        • Wennberg R.
        • Cheyne D.
        EEG source imaging of anterior temporal lobe spikes: validity and reliability.
        Clin Neurophysiol. 2014; 125: 886-902https://doi.org/10.1016/j.clinph.2013.09.042
        • Wieser H.G.
        • Blume W.T.
        • Fish D.
        • Goldensohn E.
        • Hufnagel A.
        • King D.
        • et al.
        Proposal for a new classification of outcome with respect to epileptic seizures following epilepsy surgery.
        ILAE Comm Rep. 2001; 42 (Epilepsia): 282-286
        • Zijlmans M.
        • Zweiphenning W.
        • van Klink N.
        Changing concepts in presurgical assessment for epilepsy surgery.
        Nat Rev Neurol. 2019; 15: 594-606https://doi.org/10.1038/s41582-019-0224-y